C. Operating current:30mA
D. Resolution: SpO
2
1%
Pulse Rate 1bpm
E. Accuracy: SpO
2
70% to 100%±2 digits, below 70% unspecified
Pulse rate:±2 bpm or ±2% (select larger)
F. Measurement at low perfusion:
The values of SpO
2
and pulse rate can be displayed properly when pulse saturation is at 0.4%.
Accuracy of SpO
2
:±4%
Accuracy of Pulse rate:±2bpm or ±2% (select larger)
G. Resistance to surrounding light:
The difference between the value measured in the condition of man-made light and indoor natural
light and that of darkroom is less than ±1%.
H. Automatic shutdown function:
The device will power off automatically within 5 seconds when there is no finger in the device.
5. Accessories
A. A hanging rope
B. Two batteries
C. A User Manual