The unit will not
power on
Are the batteries exhausted? Replace the batteries.
Are the batteries installed correctly? Insert the batteries observing polarity.
Stimulation weak
or cannot feel any
stimulation
Electrodes are dried out or dirty. Replace with new electrodes.
Electrodes do not stick to skin well. Replace with new electrodes.
Stimulation is
uncomfortable
Intensity is too high Decrease intensity.
Electrodes are too close together. Reposition electrodes to be at least 2
inches apart
Electrode active area size is too small.
Replace electrodes with ones that
no less than 2 in
Is the device being operated according to
the manual?
Please check the manual before use.
32
TROUBLESHOOTING
PROBLEM POSSIBLE CAUSES POSSIBLE SOLUTION