EasyManua.ls Logo

AccuRelief ACRL-9100 - EMC: Electromagnetic Immunity Declaration

AccuRelief ACRL-9100
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
36
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The ACRL-9100 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
RF emission
CISPR 11
Class B
Class A
Complies
Guidance and manufacture’s declaration – electromagnetic immunity
The ACRL-9100 is intended for use in the electromagnetic environment specified below.
The user of ACRL-9100 should assure that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be
at least 30%.
Immunity test
IEC 60601 test level
Compliance level

Other manuals for AccuRelief ACRL-9100