Tilt Reset Error
Tilt Encode Error
Tilt Encode Not Find
Tilt Encode Disable
Check whether the encoder on the tilt is damaged.
Check whether the lead connecting the encoder on the tilt and the PCB
board is in poor contact or disconnected.
Check whether the encoder on the tilt is damaged.
Check whether the lead connecting the encoder on the tilt and the PCB
board is in poor contact or disconnected.
Check whether the position of the tilt where the magnet is installed
falls off or is damaged.
Check whether there are obstacles in the tilt operating range.
Check whether the Hall element on the tilt is damaged.
Check whether the lead connecting the Hall element on the tilt and
the PCB board is in poor contact or disconnected.
Check whether the motor on the tilt is damaged.
Check whether the related circuit of the motor drive board on the tilt
is damage.