Questions and Answers

  • V
    Victoria WhiteJul 26, 2025
    How to fix Across International AccuTemp-19 Oven that cannot hold vacuum properly?
    • W
      Wendy MartinezJul 26, 2025
      To troubleshoot vacuum issues with your Across International Oven, verify the following: * Ensure the vacuum port is open, the vent port is closed, and the pump is correctly connected to the vacuum port. * Inspect the door's sealing gasket for cleanliness, proper seating, splits, cracks, flattened or melted spots, or any other defects. * Check the door alignment to ensure it makes good contact with the gasket all around when closed. If necessary, apply a small amount of vacuum grease to the door gasket to check the contact. Realign the door if the grease doesn't transfer evenly to the glass when opened.
  • J
    john65Aug 2, 2025
    Why does my Across International AccuTemp-19 Oven temperature keep overshooting?
    • P
      Philip WuAug 2, 2025
      It's normal for Across International Oven temperatures to overshoot slightly before stabilizing, typically within 2 to 5 degrees. However, if the oven consistently overheats, consider the following: * Ensure the temperature controller isn't set much higher than the calibrated setting. * Always operate the oven under vacuum. * Recalibrate the oven using Auto Tune.
  • L
    linjoeAug 8, 2025
    What to do if Across International AccuTemp-19 vacuum gauge cannot reach 30 inches of mercury?
    • A
      Aaron RaymondAug 9, 2025
      If your Across International Oven vacuum gauge struggles to reach 30 inches of mercury, verify that the vacuum pump is appropriately sized for your oven and receives adequate power. Also, check for significant off-gassing from samples and ensure the door is properly sealed. Your altitude relative to sea level also plays a role.
  • T
    Tracey BradfordAug 17, 2025
    Why is there a difference between displayed and reference temperature in Across International Oven?
    • O
      ochandlerAug 17, 2025
      When verifying temperature in your Across International Oven with an independent thermocouple, remember that temperature stabilization takes time due to the absence of air. Shelves may be cooler than the walls during warm-up. Here’s what to consider: * Always maintain a vacuum. * Allow 1 to 4 hours for stabilization. * Attach probes to shelves with heat-resistant tape. * Readings should be within the uniformity tolerance specified. If shelves are too cool, increase the temperature setting until the parts on the shelves reach the desired temperature. You can also manually offset temperature differences in the advanced settings using the “Calibration by Formula” process.

Summary

PRECAUTIONS

2. SPECIFICATIONS

3. OVEN COMPONENTS

4. PREPARATION

5. OVEN OPERATION

6. AUTO TUNE (AT) CALIBRATION

7. ADVANCED SETTINGS (FOR 1-SEGMENT CONTROLLER)

8. MULTI-SEGMENT RAMP TEMPERATURE CONTROLLER OPERATION

9. TROUBLESHOOTING

9.1. Issue: Oven temperature is consistently overshooting

Solutions for oven temperature overshoot during operation.

9.2. Issue: Temperature displayed and reference value are different

Troubleshooting steps for discrepancies between display and measured temps.

9.3. Issue: Oven cannot hold vacuum properly

Diagnosing and fixing issues with the oven's vacuum retention.

9.4. Issue: Oven vacuum gauge cannot get to ultimate 30 inches of mercury vacuum

Addressing problems with achieving the maximum vacuum level.

10. MAINTENANCE

Across International AccuTemp-19 Specifications

General IconGeneral
BrandAcross International
ModelAccuTemp-19
CategoryOven
LanguageEnglish

Related product manuals