16
Guidance and manufacture’s declaration - RF wireless communication equipment immunity
Pulse
modulation
b)
18 Hz
FMc)
5 kHz standard
deviation
1 kHz sinusoidal
710
745
780
810
870
930
1720
1845
1970
5240
5500
5785
2450
Pulse
modulation
b)
217 Hz
Pulse
modulation
b)
217 Hz
Pulse
modulation
b)
217 Hz
Local network
802.11
year wireless
Pulse
modulation
b)
217 Hz
Pulse
modulation
b)
18 Hz
GSM 800/900;
TETRA 800;
IDEN 820;
CDMA 850;
Band LTE 5
Bluetooth,
Wireless LAN, 802.11
b / g / n, RFID 2450,
LTE Band 7
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
Band LTE 1, 3, 4,
25; UMTS
a) Some services include only uplink frequencies.
b) The carrier wave should be modulated using a 50% duty cycle square wave modulated.
c) As an alternative to frequency modulation, 50% pulse modulation at 18 Hz can be used because, although it does not
represent actual modulation, it would be the worst option.
1,8
2
0,2
2
2
2
0,2
380-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
385
450
TETRA 400
GMRS 460,
FRS 460
Band LTE 13, 17
0,3
0,3
0,3
0,3
0,3
0,3
0,3
27
28
9
28
28
28
9
Frequency
Test (MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
Guidance and manufacture’s declaration - electromagnetic emission
The Infrared Body Thermometer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Infrared Body Thermometer should assure that it is used in such an environment.
Emission test Compliance
RF emissions CISPR 11 Group 1
RF emission CISPR 11 Class B
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage Fluctuations/flicker emissions
Not applicable
IEC 61000-3-3