6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.5 Measurement Function
5-28
5.2.5 Measurement Function
5.2.5.1 Measurement Function
The following measurement functions are available.
1. Voltage measurement function
2. Current measurement function
3. Resistance measurement function
For the voltage source function, resistance values are displayed by measuring current.
For the current source function, resistance values are displayed by measuring voltage.
For more information on resistance value calculation for the pulse source mode, refer to Section
5.2.5.3 , "Measurement Delay Time and Measurement Value."
When the measurement function link mode is set to OFF, the measurement function can be selected inde-
pendently from the source function. When it is set to ON, the measurement function is set according to the
source function.
Use E) MEASURE and 6) Mfunc Link to switch this mode ON or OFF.
When this mode is set to ON, the measurement function is set as follows by changing the source function.
• Voltage source function/current measurement function VSIM
• Current source function/voltage measurement function ISVM
Also, immediately after the measurement function link mode is set to ON, the measurement function
changes according to the currently set source function as shown above.
NOTE: The resistance measurement function shows the following messages if it cannot normally execute
resistance value calculation.
Count Few: The current source value is below 20 digits, or current measurement value is
below 200 digits.
VSource=0: The voltage source value is 0.
HiLimit RM: HI limiter status
Lo Limit RM: LO limiter status