6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.7 Sweep Measurement
2-43
2.2.7 Sweep Measurement
This section describes how to read out the measurement data from the memory by using the sweep source
mode.
In voltage source current measurement (VSIM), linear sweep is performed with steps of 1 V from 1 V to
10 V as shown in Figure 2-17 below.
Figure 2-17 Sweep Measurement
Source mode: DC sweep
Sweep type: Linear sweep (default)
Bias value: 0 V (default)
Start value: 1 V
Stop value: 10 V
Step value: 1 V
Integration time: 1 PLC (default)
Source delay time (Tds): 1 ms
Measurement delay time (Td): 4 ms (default)
Period time (Tp): 100 ms
Current limit value: 30 mA
Measurement data memory address
Bias value
Start value
Stop value