6247G/6247C DC Voltage Current Source/Monitor Operation Manual
2.2.7 Sweep Measurement
2-45
12. Press 123…, 1, 0 and ENTER in order.
The stop value is set to 10 V.
13. Press and to select 3) Step Value. Press to enter the Input/Run
level.
14. Press 123..., 1, and ENTER in order.
The step value is set to 1 V.
15. Press and to select 6) Bias Value.Verify that the bias value is set to 0 V.
Setting the sweep time
16. Press and to select D) TIME.
17. Press and to select 2) Src Delay. Press to enter the Input/Run level.
18. Press 123…, 1 and ENTER in order.
The source delay time is set to 1 ms.
19. Press and to select 3) Meas Delay. Verify that the measurement delay
time is set to 4 ms.
20. Press twice to select 5) Period. Press to enter the Input/Run level.
21. Press 123…, 1, 0, 0, and ENTER in order.
The period time is set to 100 ms.
Setting the measurement data memory
22. Press , , and in this order to select F) MEMORY.
23. Press to select 1) Store Mode. Press to enter the Input/Run level.
24. Rotate to select Normal.
The measurement data memory is set to Normal-ON.