19
Sector pointer area: two pointers to show the high/low alarm limit pressure percentage, the 3 pointer shows the current
rd
pressure .percentage
Scale bar graph: (0%~100%) FS, minimum scale is 2%FS.
Scale midpoint: point to 50%.
Example: for the ADT681 ,if the current pressure reading is 50.00kPa (50%FS) and(pressure range 0-100kPa)
the and the lower limit is 40kPa (40%FS), the analogue dial showhigher limit is 80kPa (80%FS) will
as figure 7-7-3.
Figure 7-7-3: Overpressure Alarm
When the current pressure is beyond the range of the hi low limit, the whole screen will flash to warn the user to/gh
adjust the pressure. Meanwhile, In order to catch up to the pressure change, the measure speed will automatically
adjust to ten times per second will. Once the alarm is over, the display go back to normal speed.
7.7.4 Setting alarm limit
To set the alarm limit:
(1) Press and hold display the higher limit and the lower limit then move the cursor up or down by
.pressing or