Function Introduction
Right return value
A X Knnnn
C0 C1 C2 C3 C4
Eos
W
OFRUN
---
0x0
Set data logging start or stop
A:F: OFRUN:OK+Eos
0(stop)
1(start)
W
OFTIM
----
interval
Set recording interval (1s~99999s)
A:F: OFTIM:OK+Eos
R
OFSTA
211
--
0x0
Read the operating status of data logging
A:F: OFSTA:state:interval:space:total records+Eos
OFDEL
-
----
0x0
Delete all data records
A:F: OFDEL:OK+Eos
W
OFSAP
---
0x0
Send all data records
R
Read RTC
A:F: OFSAP:OK+Eos
Automatically send data packets, detailed packet
format, please contact the manufacturer
W
ORTC
yymmddhhmmss
----
0x0
Write RTC
A:F:ORTC: yymmddhhmmss +Eos
0x0
W
ORTC
---
0x0
A:F:ORTC:OK+Eos
-
-
41
0(stop)
1(start)
-
--
-
W
OCPS
W
OCP
W
OCPOK
WALARM
RALARM
WMRATE
RMRATE
W
ODIAL
W
ORPP
-
Z (zero)
M(middle)
F (full scale point)
1( )save
()0 unsaved
High limit
-
D0
0(%) 1(swing)
2( larm)a
-
-
----
0x0
Entrance instruction of calibration
A F OCPS OK+Eos:: :
---
0x0
Input the standard pressure value
and calibration points, for calibration
A F OCP OK+Eos:: :
Standard
pressure of
calibration
point
lower
limit
----
Quit the pressure calibration process
A F OCPOK OK+Eos:: :
--
0x0
Set the alarm limit
A F ALARM OK+Eos:: :
Pressure
unit
----
0x0
Read alarm limit
A F ALARM : ::: :high limit lower limit
pressure unit+Eos
---
D1
0x0
Set measure D1 times/ D0 second A F MRATE OK+Eos:: :
0x0
Read the measured times per second
AFMRATE:: :seconds: times+Eos
----
0x0
Set the work mode on the analog dial
A F ODIAL OK+Eos:: :
----
0x0
---
0x0
Software Reset
A F OK+Eos:: :ORPP
-