EasyManua.ls Logo

ADE BE 1817 - Cleaning; Fault;Remedy

ADE BE 1817
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning
19
Digital baby scale BE 1817
English
Cleaning
NOTE on material damage
Do not immerse the scales in water or other liquids.
Do not under any circumstances use harsh, scraping, or abrasive
cleaning agents or the like to clean the scales. If you do, the surface
could be scratched.
Wipe down the scales as needed using a lightly dampened, soft cloth.
Disinfect the weighing surface from time to time using a suitable disinfecting
agent (available at specialist retailers and pharmacies).
Fault/Remedy
Fault Possible Cause
No function Battery empty or inserted the wrong way
around?
The scales appear to be displaying
false weighing results
Are the scales standing on a rm, even
surface?
Is a device that can cause electromagnetic
disturbances located in the vicinity of the
scales?
Remove the battery from the scales and wait
a moment. Reinsert the battery.
“O-Ld” is shown on the display The maximum capacity of the scales (20 kg)
has been exceeded.
“LO” is shown on the display The batteries are empty
(see “Inserting/changing batteries”).
“UnSt” is shown on the display The scale was unstable during the
measurement. Check the base of the scale
and repeat the weighing.
“- - - -” is shown on the display The scale is not ready or is positioned
incorrectly. Reposition the scale.