Electromagnetic Compatibility
Like other electrical medical equipment, the AMM213TD/AMM240WTD requires special
precautions to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the AMM213TD/AMM240WTD must be installed
and operated according to the EMC information provided in this manual.
Note The AMM213TD/AMM240WTD has been designed and tested to comply with
IEC 60601-1-2:2001 requirements for EMC with other devices.
Caution Portable and mobile RF communications equipment may affect
the normal function of the AMM213TD/ AMM240WTD.
Warning Do not use cables or accessories other than those provided
with the AMM213TD/AMM240WTD, as this may result in increased electromag-
netic emissions or decreased immunity to such emissions.
Warning If the AMM213TD/AMM240WTD is used adjacent to or stacked
with other equipment, observe and verify normal operation of the AMM213TD/
AMM240WTD in the conguration in which it will be used prior to using it in
a
surgical procedure. Consult the tables below for guidance in placing the AM-
The Model AMM213TD/AMM240WTD is intended for use in the electromagnetic environment specied below.
The customer or the user of AMM213TD/AMM240WTD should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1
The Model AMM213TD/AMM240WTD uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic
equipment
RF emissions CISPR 11 Class B
AMM213TD/AMM240WTD is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes
Harmonic emissions
IEC61000-3-2
Class D
Voltage Fluctuations
IEC61000-3-3
Complies
Manufacturer’s declaration - electromagnetic emission
The Model AMM213TD/AMM240WTD is intended for use in the electromagnetic environment specied below.
The customer or the user of AMM213TD/AMM240WTD should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient
/burst IEC61000-4-4
2 kV for power supply lines
1 kV for input / output lines
2 kV for power supply lines
1 kV for input / output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Manufacturer’s declaration - electromagnetic immunity