E
2
T QUASAR Flare Monitors Operation Manual Appendix A: Specifications and Parameter
Settings 56
Hazard Classification
(See Appendix B, Area
Classifications / Protection
Concepts.)
Explosion Proof, prevent flame propagation
CSA/US
Class 1, Division 1, Groups C and D
Class 1, Division 2, Groups A through D
Temp code T4A
Enclosure Type 4X
CENELEC/ATEX approved Zone 1
Type : Ex db IIB T4
Threads
Electrical service: ¾ inch NPT conduit
cooling base service: ¼ inch NPT
Mounting
M-3 mounting fixture height adjustable pedestal with 360 ° swivel
70 ° to +70 ° vertical
Torque Spec, Lid Bolts
50in-lbs, + / -5; 5.5N-m
Air
View port purge 20psig min., 1 SCFM
Cooling (For ambient
environments over 120F)
WATER: 10 gph (37.8 lph)
AIR: Vortex, V208-15H, 10SCFM minimum
Reference manufacturer data for pressure specs.
Weight/Dimensions
48 Pounds (22 Kg) / 5 x 12.5 x 8.5 in (127 x 318 x 217 mm)
Varying Flame
Luminescence
The LumaSense E
2
T
QUASAR M8100-EXP automatically compensates
for low and varying target emissivity (luminescence) and minimizes
the effects of attenuation by intervening media, such as rain, snow
and fog.