Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The model PDC-W240 is intended for use in the electromagnetic environment specified
below. The customer or the user of the model PDC-W240 should assure that it is used
in such an environment.
Immunity
IEC 60601 Test
Compliance
Electromagnetic Environmental
test
Level
Level
Guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the model
PDC-W240, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Dis-
tance
d = 1,2
P
d = 1,2
P 80 MHz to 800 MHz
Conducted
3 Vrms
RF
d = 2,3
P 800 MHz to 2,5 GHz
150 kHz to 80
IEC 61000-4-
MHz
Vrms
6
where P is the maximum output
Radiated RF
3 V/m
V/m
power rating of the transmitter in
watts (W) according to the transmit-
80 MHz to 2,5
IEC 61000-4-
ter manufacturer and d is the rec-
GHz
3
ommended separation distance in
meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey,
a
should be
less than the compliance level in
each frequency range.
b
Interference may occur in the vicin-
ity of equipment marked with the fol-
lowing symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the PDC-W240 is used
exceeds the applicable RF compliance level above, the PDC-W240 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advantech PDC-W240 and is the answer not in the manual?

Advantech PDC-W240 Specifications

General IconGeneral
BrandAdvantech
ModelPDC-W240
CategoryMonitor
LanguageEnglish