MODEL 3711/3731 SERVICE/CALIBRATION MANUAL Cat. #2117.60/2117.61
v3 – 03/02 - 6 -
CALIBRATION EQUIPMENT
REQUIRED ITEMS:
For Field Calibration: (resistance values only)
Each ground resistance tester is supplied with a calibration loop to check and calibrate it. This loop
does not allow calibration above 198Ω. The typical range of measurement of ground resistance is less
than this value.
One DC power supply adjustable from 5.6V to 10V (0.2A min.)
Four resistors:
20Ω, 120Ω, 150Ω, 500Ω (Accuracy 0.25%)
One AC Current generator to supply 1.4A, 50 or 60Hz (Accuracy 0.25%)
One small insulated flat screwdriver.
For Laboratory Calibration: (including amps range):
Add the following items:
One decade resistance box, accuracy 0.25% (non-inductive) adjustable from 0.01Ω to 500Ω.
One AC current generator capable of supplying 100mA to 20A into a two turn wire loop with a
frequency of 50 to 400Hz. (Fluke 5500 or 5520).
One DMM 20000 cts (Accuracy: < 0.1% in DC) and (< 0.5% in AC).