This quick start guide has been developed by the NHS and Cabinet Office and should be used in conjunction with the manufacturers Instructions for Use
CONNECT AND SET UP DEVICE
What controls does the ventilator have? (Image E)
3
SET THE CONTROLS ON THE VG70
AEONMED VG70 VENTILATOR: QUICK START GUIDE
CONNECT AND SET UP THE VG70
CLASS A ICU
Check oxygen sensor is
connected under flap below
inspiratory port
Attach HMEF to expiratory port (D)
Connect patient circuit to
inspiratory port and to HMEF on
expiratory port.
A standard humidification setup is
possible as shown
The Y piece is then connected to a
test lung and then patient.
The device is largely touch screen menu driven (image E)
with a control/confirm knob to the side of the screen
adjacent to an alarm mute button.
The controls are accessed by pressing the main menu
button (highlighted yellow in the image)
The standby button/start ventilation is in the bottom right
hand corner of the screen (highlighted blue in the image)
These controls are used to input patient data and select the
standard modes of ICU ventilation and also to adjust the
parameters, e.g. Tidal volume, Respiratory rate, I:E, Fi02,
PEEP/CPAP etc, as well as to set limits.
Expiration valve
lock
Inspiration port
Expiration valve
Expiration port
Oxygen sensor
storage shell
Nebulizer
Connect mains power adapter to the power plug
Connect Oxygen from the wall or cylinder to the
oxygen input port.
Air connection is not required as this ventilator is
turbine driven
How do I connect the ventilator?
(Image A)
1
How do I set up the breathing system
and the HME filter? (Image B)
2
A
Power
Switch
Oxygen
input port
B
C
D
E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Aeonmed vg70 and is the answer not in the manual?

Aeonmed vg70 Specifications

General IconGeneral
TypeICU Ventilator
Tidal Volume20 to 2000 ml
PEEP0 to 50 cmH2O
Pressure Support0 to 60 cmH2O
WaveformsPressure, Flow, Volume
Modes of VentilationVCV, PCV, PSV, SIMV, APRV
Respiratory Rate1 to 100 bpm
Inspiratory Time0.2 to 5 seconds
Inspiratory to Expiratory Ratio1:1 to 1:10
FiO221% to 100%
Trigger Sensitivity (Flow)1-10 L/min
Display12.1 inch TFT touch screen
Power SupplyAC 100-240V, 50/60Hz