C-22 Subject to Export Control, see Cover Page for details.
C-4. REMOTE OPERATION COMMANDS (cont)
External Audio Output
COMMAND RANGE DESCRIPTION
:extaudout:source 0 - EXT_AUD_IN_2_EXT_
AUD_OUT
1 - DEMOD_2_EXT_AUD_
OUT
2 - MODULATION_2_EXT_
AUD_OUT
3 - FGEN_2_EXT_AUD_OUT
Selects signal source.
:extaudout:source? Returns external audio output source.
:extaudout:state ON or OFF Sets external audio output ON/OFF.
:extaudout:state? Returns external audio output state.
Frequency Find
COMMAND RANGE DESCRIPTION
:freqfind:peak? in MHz Returns next tune frequency.
:freqfind:start 2 to 1000 MHz Sets tune start frequency.
:freqfind:start? Returns tune start frequency.
:freqfind:stop 2 to 1000 MHz Sets tune stop frequency.
:freqfind:stop? Returns tune stop frequency.
:freqfind:threshold -110.0 to 40.0 dBm Sets tune threshold.
:freqfind:threshold? Returns tune threshold.
:freqfind:channel_boundary 0.001 to 5.000 MHz Sets channel bandwidth step for frequency
search.
:freqfind:channel_boundary? Returns channel bandwidth step for frequency
search.
Function Generator
COMMAND RANGE DESCRIPTION
:fgen:enable Sets Function Generator ON/OFF.
:fgen:enable? Returns Function Generator condition.
:fgen:freq 1 - Fgen1
2 - Fgen2
0 to 24000 Hz
Sets individual Function Generator frequency.
:fgen:level 1 - Fgen1
2 - Fgen2
0 to 1.7 Vrms
Sets individual Function Generator output level.
:fgen:load
0 - 600 Ω
1 - 150 Ω
2 - Open Circuit
Sets output scaling.
:fgen:state 1 - Fgen1
2 - Fgen2
ON or OFF