Aeroneb
®
Professional Nebulizer System Instruction Manual 13
Assembly and Installation
1. Clean and sterilize the nebulizer unit and T-piece(s) as
described in the Cleaning, disinfection and sterilization
section of this manual.
Note: The nebulizer unit and T-piece, as packaged, are not
sterile.
2. Perform a functional test of Aeroneb Pro before use and
between patients as described in the functional test
section of this manual.
3. Insert the filler cap into the opening on the nebulizer unit.
4. Connect the nebulizer unit to the T-piece by pushing the
nebulizer unit firmly onto the T-piece (Figure 3).
Figure 3: Connecting nebulizer unit to T-piece