7. Power on and observe Aerogen Ultra to confirm aerosol is visible.
8. Introduce Aerogen Ultra to patient and observe aerosol flow to ensure correct operation.
9. Remove excess rainout from the Aerogen Ultra periodically (hourly with continuous nebulization).
To ensure optimum performance of the Aerogen Ultra, remove any residue by rinsing through
with sterile water, shake off excess and allow to air dry.
Warnings
Performance
Do not use with a closed face mask or a standard oxygen mask.
When using with an aerosol face mask, always use supplemental oxygen flow of 1-6 LPM for adult
and a maximum of 2 LPM for pediatric patients less than 18 years of age.
Performance of the Aerogen Ultra may vary depending upon the type of drug and Aerogen Ultra
configuration used.
Do not exceed recommended oxygen flow for system.
Visually check Aerogen Ultra post-rinsing to ensure that valves have not become dislodged.
When connecting a 22mm(M) breathing system filter to the Aerogen Ultra Mouthpiece, ensure that
the filter does not occlude the exhalation valve of the mouthpiece.
If using a non-Aerogen supplied filter, change filter as per manufacturer instructions or more
frequently if it becomes obstructed.
The Aerogen Ultra, when used in combination with the Mouthpiece, Aerogen Solo nebulizer and a
filter (with a recommended minimum efficiency rating of 99.9% (Bacterial) or 99.8% (Viral)) has the
ability to reduce but not eliminate the risk of transmission or acquisition of an infectious agent by
healthcare workers or others.
The Aerogen supplied filter must be dry and free of secretions. Wet filter material may result in low
filtration efficiency.
Do not attempt to sterilize, autoclave or clean the Aerogen supplied filter.
General
To be used by trained personnel only.
Use only with recommended components.
Ensure tubing is safely orientated to prevent strangulation hazard.
Ensure oxygen connection port or tubing is not occluded.
Do not use the Aerogen Ultra without a mouthpiece or face mask.
Do not autoclave any component of the kit.
Do not cover Aerogen Ultra valves during use.
Aerogen Ultra Kits
AG-AS7500-US:
AG-AS7000-US:
AG-AS7501-US:
AG-AS7502-US:
AG-AS7001-US:
AG-AS7002-US:
AG-AS7509-US:
AG-AS7009-US:
AerogenUltra Starter Kit (10 pack): AerogenSolo, AerogenUltra, oxygen tubing
AerogenUltra (10 pack): AerogenUltra, oxygen tubing
Aerogen Ultra Starter Kit - Mask (10 pack): Aerogen Solo, Aerogen Ultra, oxygen
tubing, I-Guard Aerosol Mask (Adult)
Aerogen Ultra Starter Kit - Mask (10 pack): Aerogen Solo, Aerogen Ultra, oxygen
tubing, I-Guard Aerosol Mask (Pediatric)
Aerogen Ultra - Mask (10 pack): Aerogen Ultra, oxygen tubing, I-Guard Aerosol
Mask (Adult)
Aerogen Ultra - Mask (10 pack): Aerogen Ultra, oxygen tubing, I-Guard Aerosol
Mask (Pediatric)
Aerogen Ultra Starter Kit with filter (10 pack): Aerogen Solo, Aerogen Ultra, oxygen
tubing, filter
Aerogen Ultra with filter (10 pack): Aerogen Ultra, oxygen tubing, filter
Aerogen Supplied Disposable Bacterial/Viral Filter:
22mm ID x 15mm ID/22mm OD
Indications for Use
The disposable Bacterial/Viral filter is intended for use within respiratory devices where filtration of
particulate matter is desired. Each Aerogen supplied filter is intended for single patient use up to 24
hours, and is not intended for reuse.
Hydrophobic BFE: > 99.99%
Resistance to Flow: < 2cm H
2
O @60LPM
Hydrophobic VFE: > 99.99%
Resistance to Flow: 1cm H
2
O @30LPM
Symbols related to the Aerogen Supplied Filter
Symbol Meaning Symbol Meaning
Single patient use only
Not made with
natural rubber latex
Keep dry
Rx Only
Caution: Federal law restricts
this device to sale by or on
the order of a physician
Caution
Attention: Consult
accompanying documents
for important cautionary
information
This product is supplied non-
sterile