System Instruction Manual
www.aerogen.com

Questions and Answers

  • P
    phardyAug 16, 2025
    What to do if medication is left in the nebulizer after nebulization cycle on my Aerogen Pro?
    • C
      Christine MorrisAug 16, 2025
      If medication remains in the nebulizer after the nebulization cycle of your Aerogen Medical Equipment, it could be due to several reasons. Ensure the nebulizer is connected to power and turned on. The rechargeable battery may be depleted, so recharge it. The nebulizer may not have been properly cleaned, so clean it. If you selected a 15 Minute cycle and added a volume greater than 3 mL to the nebulizer, run an additional 15 Minute cycle, or select the 30 Minute cycle when delivering a dose greater than 3 mL.
  • J
    jenniferschmidtAug 19, 2025
    What to do if the 15 Min. or 30 Min. indicator does not light on my Aerogen Pro?
    • S
      Sean ParkerSep 12, 2025
      If the 15 Min. or 30 Min. indicator does not light when you press the On/Off power button on your Aerogen Medical Equipment, there could be a couple of reasons. First, check that the AC/DC adapter is securely attached to the controller to ensure the system is getting power. Second, the rechargeable battery might be depleted, so try recharging it.
  • A
    Amy BrownAug 22, 2025
    Why is the 15 Min. or 30 Min. indicator flashing on my Aerogen Medical Equipment during nebulization?
    • Q
      qfergusonAug 22, 2025
      If the 15 Min. or 30 Min. indicator flashes during nebulization with your Aerogen Medical Equipment, it means the battery power is low. You should recharge the battery.
  • G
    gileserinAug 25, 2025
    Why is the treatment time longer than expected with my Aerogen Pro Medical Equipment?
    • K
      Kathy ThompsonAug 26, 2025
      If your Aerogen Medical Equipment is taking longer than expected to nebulize (for example, more than 15 minutes for 3 mL of Normal Saline), it could be due to a few reasons. The rechargeable battery might be depleted, so try recharging it. Also, the nebulizer may not have been properly cleaned, so ensure you clean it.
  • H
    Heather GarciaAug 27, 2025
    What to do if the 15 Min. or 30 Min. light is on, but aerosol is not visible from my Aerogen Pro Medical Equipment?
    • L
      lloydleonardAug 28, 2025
      If the 15 Min. or 30 Min. light is on but you don't see any aerosol coming from your Aerogen Medical Equipment, there might be a few issues. First, check if there is medication in the nebulizer and refill it through the filler cap if it's empty. Second, the nebulizer may not have been cleaned properly, so clean it.
  • C
    casejodyAug 31, 2025
    What to do if medication is left in the Aerogen Pro Medical Equipment nebuliser after nebulisation?
    • J
      Joshua EllisAug 31, 2025
      If medication remains in the nebuliser after a cycle, ensure the nebuliser is connected to power and turned on. If the rechargeable battery is depleted, recharge it. Clean the nebuliser properly. If you selected a 15-minute cycle and added a volume greater than 3 mL to the nebuliser, run an additional 15-minute cycle, or select the 30-minute cycle when delivering a dose greater than 3 mL.
  • M
    Mrs. Katherine MorenoSep 3, 2025
    Why is my Aerogen Medical Equipment treatment taking longer than expected?
    • A
      Adam StephensSep 3, 2025
      Longer than expected treatment times can occur if the rechargeable battery is depleted, so try recharging the battery. Also, ensure the nebuliser has been properly cleaned.
  • B
    Brenda WalshSep 6, 2025
    What to do if the Aerogen Pro 15 Min. or 30 Min. indicator does not light up?
    • M
      Melissa LawrenceSep 6, 2025
      If the 15 Min. or 30 Min. indicator doesn't light when you press the On/Off button, check that the AC/DC adapter is securely attached to the controller. Also, the rechargeable battery may be depleted, so try recharging it.
  • J
    Jeremy MartinezSep 9, 2025
    Why is the 15 Min. or 30 Min. light on my Aerogen Pro Medical Equipment illuminated, but there is no visible aerosol?
    • R
      Roger DrakeSep 9, 2025
      If the 15 Min. or 30 Min. light is illuminated but no aerosol is visible, first check if there is medication in the nebuliser and refill it through the filler cap. Also, a likely cause is that the nebuliser hasn't been cleaned properly, so ensure you clean the nebuliser.
  • V
    Valerie HaleSep 12, 2025
    Why is the 15 Min. or 30 Min. indicator flashing on my Aerogen Pro?
    • E
      Erin MillerSep 12, 2025
      The 15 Min. or 30 Min. indicator flashes during nebulisation because the battery power is low. Recharge the battery.

Summary

Aerogen Pro System Overview

System Components

Lists and illustrates the main components of the Aerogen Pro System, including nebulizer, T-piece, controller, cable, adapter, and bracket.

System Warnings

Aerogen Pro Controller Interface

Controls and Indicators

Explains the function of each control and indicator on the Aerogen Pro controller, such as power, timer, and status lights.

Recharging the Battery

Assembly and Installation

System Set-Up Procedure

Guides the user through cleaning, sterilization, and assembly of the Aerogen Pro System components.

Installation for Ventilator Use

Connection to Breathing Circuit

Details how to connect the nebulizer to adult and pediatric breathing circuits for ventilator use.

Installation for Ventilator Use (Continued)

Ventilator Use Warnings

Lists specific warnings for nebulizer orientation and circuit connections during ventilator use.

Adding Medication

Nebulization Operation

Off-Ventilator Installation

Off-Ventilator Installation (Continued)

Functional Test Procedure

Cleaning, Disinfection, and Sterilization

Component Handling

Lists components and stresses the importance of pre-use cleaning and sterilization for hygiene.

Cleaning, Disinfection, and Sterilization (Continued)

Cleaning, Disinfection, and Sterilization (Continued)

Cleaning, Disinfection, and Sterilization (Continued)

Nebulizer Sterilization Methods

Explains steam sterilization and hydrogen peroxide gas plasma sterilization for the nebulizer.

Cleaning the Aerogen Pro Controller

Troubleshooting Guide

Troubleshooting Guide (Continued)

Performance Specifications

Appendix 1: Electromagnetic Compatibility (Continued)

Aerogen Pro Specifications

General IconGeneral
TechnologyVibrating Mesh
Residual Volume<0.1 mL
Operating Temperature10°C to 40°C
Nebulization Rate0.5 mL/min
Medication Capacity6 mL
Operating ModesContinuous
Power SourceRechargeable Battery, AC Adapter
Power Supply100-240 V, 50/60 Hz