Cleaning the Aerogen Pro Controller
Cleaning of controller, controller cable & AC/DC adapter
1. Wipe clean with an alcohol based disinfectant wipe or a quaternary
ammonium compound based disinfectant wipe.
2. Check for exposed wiring, damaged connectors, or other defects and
replace if any are visible.
3. Visually inspect for damage and replace the controller if any damage
is observed.
Warnings
• Do not autoclave.
• Do not use abrasive or sharp tools.
• Do not spray liquid directly onto the controller.
• Do not immerse controller in liquid.
Note: The Aerogen Pro nebulizer contains active electronic components.
Aerogen has validated the methods of cleaning, disinfection and
sterilization above. The use of any other means of cleaning, disinfection
or sterilization has not been validated and is likely to reduce the life of your
nebulizer and will invalidate your warranty.
Cleaning of mounting brackets
Wipe clean with an alcohol based disinfectant wipe or a quaternary
ammonium compound based disinfectant wipe. Do not use abrasive or
sharp tools.