Aerogen
®
Pro System Instruction Manual
9
Assembly & Installation
Aerogen Pro System Set-Up
Clean and sterilize the nebulizer and T-piece(s) as described in the
Cleaning, Disinfection and Sterilization section of this manual.
Note: The nebulizer and T-piece, as packaged, are not sterile.
Perform a functional test of Aerogen Pro before use and between
patients as described in the Functional Test section of this manual
(see page 19).
Insert the filler cap into the opening on the nebulizer.
Connect the nebulizer to the T-piece by pushing the nebulizer firmly
onto the T-piece (Figure 3).
Figure 3. Connecting nebulizer to T-piece