Aerogen
®
Pro System Instruction Manual
29
Specifications
Table 3. Physical Specifications of the Aerogen Pro System
Nebulizer Dimensions
45 mm H x 50 mm W x 50 mm D
1.8” H x 2.0” W x 2.0” D
Aerogen Pro Controller
Dimensions
33mm H x 75mm W x 131mm D
1.3” H x 2.9” W x 5.2”D
Controller Cable Length 1.8 m (5.9 ft.)
AC/DC Adapter Cable Length 2.1 m (6.7 ft.)
Nebulizer Weight 25 g (0.9 oz) nebulizer and filler cap
Aerogen Pro Controller Weight 230 g (8.1 oz.), including battery and cable
Nebulizer Capacity Maximum 10 mL
Table 4. Environmental Specifications of the Aerogen Pro System
Operating
Maintains specified performance at circuit pressures up to 90cm
H
2
O and temperatures from 5 ºC (41ºF) up to 45 ºC (113ºF).
Atmospheric Pressure 450 to 1100 hPa
Humidity 15% to 95% relative humidity
Noise Level < 35 dB measured at 0.3 m distance
Storage & Transport
Transient Temperature
Range
-20 to +60°C (-4 to +140°F)
Atmospheric Pressure 450 to 1100 hPa
Humidity 15 to 95% relative humidity
Table 5. Power Specifications of the Aerogen Pro System
Power Source
Can operate from AC/DC Adapter (input 100 to 240 VAC 50 – 60
Hz, output 9 V) or internal rechargeable battery (4.8 V nominal
output).
Note: The Aerogen Pro controller is approved for use with
Aerogen AC/DC adapter AG-AP1040-US (Manufacturer
Reference: FRIWO FW8000M/US/09 / FW8000M/09 /
FW7660M/09)
Power Consumption ≤ 6.5 Watts (charging), ≤ 2.0 Watts (nebulizing).
Patient Isolation
Controller circuitry provides 4 kilovolt (kV) patient isolation and
complies with IEC/EN 60601-1.