26
Aerogen
®
Troubleshooting
If these suggestions do not correct the problem, discontinue use of any
device and contact your local Aerogen sales representative.
Table 2. Aerogen Pro Troubleshooting
If this happens: It could mean: Try this:
The 15 Min. or 30 Min.
indicator flashes during
nebulization.
Battery power is low.
Recharge battery
(see Recharging the Battery).
Battery will not
recharge.
Constant green light
showing on the battery
status indicator and
flashing green light on
either the 15 Min. or
30 Min. indicator light,
when the controller is
connected to the AC/DC
Adapter.
It may be time to replace
the battery.
Contact your local Aerogen
sales representative.
Battery will not retain
initial charge.
Rechargeable battery may
need to be replaced.
Contact your local Aerogen
sales representative.
The 15 Min. or 30 Min.
light illuminates, but
aerosol is not visible.
No medication in nebulizer.
Refill medication through filler
cap in the nebulizer (see page
15).
Nebulizer has not been
cleaned properly.
Clean nebulizer (see page
20)
It may be time to replace
the nebulizer.
See Warranty and Life of
Product. Refer to Aerogen
Pro parts list by visiting
www.aerogen.com.
15 Min. or 30 Min.
indicator does not light
when On/Off power
button is pressed.
There is no power to the
system.
Verify that AC/DC adapter
is securely attached to
controller.
Rechargeable battery is
depleted.
Recharge battery (see
Recharging the Battery).
The error indicator light
illuminates.
The controller cable is
incorrectly connected to
the nebulizer, or electronics
are malfunctioning.
Verify that controller cable
is correctly connected to
both the nebulizer and the
controller.