Aerogen
®
Pro System Instruction Manual
27
Table 2. Aerogen Pro Troubleshooting (Continued)
If this happens: It could mean: Try this:
Longer than expected
treatment time. e.g. 3
mL of Normal Saline
(0.9%) should take no
longer than 15 minutes
to nebulize.
Rechargeable battery is
depleted.
Recharge battery (see
Recharging the Battery).
Nebulizer has not been
properly cleaned.
Clean nebulizer (see page
20).
It may be time to replace
the nebulizer.
See Warranty and Life of
Product. Refer to Aerogen
Pro parts list by visiting
www.aerogen.com.
Medication is left in
the nebulizer after
nebulization cycle.
Nebulizer was not turned
on or connected to power.
Ensure that nebulizer is
connected to power and
turned on.
Rechargeable battery is
depleted.
Recharge battery (see
Recharging the Battery).
Nebulizer has not been
properly cleaned.
Clean nebulizer (see page
20).
A 15 Minute cycle was
selected and a volume
greater than 3 mL was
added to the nebulizer.
Run an additional 15 Minute
cycle. When delivering a dose
greater than 3 mL select the 30
Minute cycle.
It may be time to replace
the nebulizer.
See Warranty and Life of
Product. Refer to Aerogen
Pro parts list by visiting
www.aerogen.com.
Note: The rechargeable battery in the Aerogen Pro controller should only
be replaced by Aerogen authorized personnel: contact your Aerogen sales