Manual Cleaning
Cleaning nebulizer, T-pieces and Adapters
1. Ensure there is no medication remaining in the device.
2. Remove nebulizer from T-piece. Remove filler cap from nebulizer.
3. Clean all parts with warm water and mild liquid detergent.
4. Rinse parts with sterile water.
5. Shake excess water from parts and allow parts to fully air dry.
Warning: Do not use abrasive or sharp tools to clean the nebulizer.
Disinfection
Aerogen Pro nebulizer, T-pieces and Adapters with disinfection agents.
1. Follow steps 1 through 3 in Manual Cleaning section.
2. Completely immerse parts in appropriate disinfecting agent in
accordance with current hospital protocols and disinfectant agent
manufacturer guidelines.
Note: Aerogen approves the following disinfection solutions for use with
its Aerogen Pro nebulization system regarding material compatibility. With
respect to microbiological effectiveness, please ask the manufacturer.
Refer to the product labeling for specific instructions regarding activation,
safe use and disposal of these solutions.
• Isopropyl (70%)
• CIDEX
®
• NU-CIDEX
®
• CIDEX
®
OPA
• Hexanios G+R
Warning: The use of any other means of cleaning, disinfection or
sterilization has not been qualified and is likely to reduce the life of your
nebulizer and will invalidate your warranty.