Appendix 1
EMC tables:
The following Tables are provided in accordance with IEC/ EN 60601-1-2:
Guidance and manufacturers declaration – electromagnetic emissions
The Aeroneb Go nebulizer system is intended for use in the electromagnetic
environment specied below. The customer or the user of the Aeroneb Go
nebulizer system should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic
environment – guidance
RF Emissions
CISPR 11
Group 1
The Aeroneb Go nebulizer
system uses RF energy
only for its internal
function. Therefore, its RF
emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class B
The Aeroneb Go nebulizer
system is suitable for
use in all establishments,
including domestic
establishments and those
directly connected to the
public low voltage power
supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not Applicable