TROUBLESHOOTING
IMPORTANT
Before attempting any troubleshooting, ensure that the battery is fully charged.
SYMPTOM POSSIBLE CAUSE TROUBLESHOOTING STEPS
AffloVest controller
will not power on.
Battery discharged
OR
Battery not connected
properly
OR
Possible damaged
controller
OR
Possible damaged
battery
Ensure battery is fully charged, attempt to operate
when connected to power supply.
Verify that the battery is connected to the AffloVest
correctly
Attempt to operate the AffloVest when connected to
the Power Supply
Contact your dealer for further assistance.
The selected
program function
has suddenly
switched off.
The therapy cycle has
finished and a new
therapy cycle must be
started.
OR
Battery discharged
Start a new therapy cycle.
Ensure battery if fully charged, attempt to operate
when connected to power supply.
Contact your dealer for further assistance.
Controller is on
but motors will not
turn on.
Battery discharged
OR
Possible damaged
controller
Ensure battery is fully charged, attempt to operate
when connected to power supply.
Contact your dealer for further assistance.
ELECTROMAGNETIC GUILDEINES FOR CLINICAL APPLICATIONS
The AffloVest is intended for use in the electromagnetic environment specified
below. The customer or user of the AffloVest should ensure that it is used in such an
environment.
EMITTED INTERFERENCE
MEASUREMENT
CONFORMITY ELECTROMAGNETIC
ENVIRONMENT
GUIDELINES
RF emissions as per CISPR 11 Class B The AffloVest is suitable for
use in all establishments,
including domestic, and those
directly connected to the
public power supply network
that supplies buildings used
for domestic purposes.
Harmonic component emissions as per IEC 61000-
3-2
Class A
Voltage fluctuations/flicker emissions as per IEC
61000-3-2
Compliant