PROCEDURE AP2.001
IMBALANCE SETTING
Notice Originale - Rev.1.4 Manuel Maintenance – Lisa / Loreena 2/ 2
Step 3: repeat the operation with a
weight of 25 grams then press on the
validation" button
Step 4: check the imbalance calibration
with 15 & 25 grams weights.
Screen capture (25 g test)
Screen capture (validation of test)