EasyManuals Logo

Agilent Technologies 1159A User Manual

Agilent Technologies 1159A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
9
1159A 1GHz Active Differential Probe
Specifications and Characteristics
The following characteristics are valid for the 1159A probe after the probe has
reached operating temperature, which is 20 minutes with power applied in a
environment with stable ambient temperature. The probe must be operating
within the environmental conditions listed in the “Environmental Specifications”
section on page 11, and must have been calibrated within the past 12 months in
a ambient temperature of 23 5 C.
Characteristics
Probe Bandwidth (-3 dB) DC to 1 GHz
Offset Range 1.6 V
Rise Time (Probe only)
1:1 Attenuation
Internal switched attenuation only
<350 ps
Input Resistance
(each side to ground)
1 M
Input Capacitance (between inputs)
1:1 Attenuation
No external attenuators
<0.85 pF
Input Capacitance (each side to ground)
1:1 Attenuation
No external attenuators
<1.5 pF
±°
±
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com