8
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Specifications and Characteristics
Specifications
Range
Input Configuration Ground Connector
True Differential (+ and inputs), with shield
Input Coupling DC AC coupling obtained by installing an AC coupling
adapter
Gain Accuracy at 1 kHz 2%
Maximum Input Voltage
Either input from ground
< ±42 V
CMRR at 70 Hz: 80dB
at 1 MHz: 40dB
at 100 MHz: 25dB
at 500 MHz: 19dB
at 1 GHz: 13dB
Ð
!
Mode No attenuator ÷10 attenuator ÷20 attenuator
Differential < ±400 mV < ±4 V < ± 8V
Common < ±16 V < ±42 V < ±42 V
Offset (Common) < ±1.6 V < ±16 V < ±32 V