EasyManua.ls Logo

Agilent Technologies 1159A - Specifications and Characteristics; Specifications; Range

Agilent Technologies 1159A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Specifications and Characteristics
Specifications
Range
Input Configuration Ground Connector
True Differential (+ and inputs), with shield
Input Coupling DC AC coupling obtained by installing an AC coupling
adapter
Gain Accuracy at 1 kHz 2%
Maximum Input Voltage
Either input from ground
< ±42 V
CMRR at 70 Hz: 80dB
at 1 MHz: 40dB
at 100 MHz: 25dB
at 500 MHz: 19dB
at 1 GHz: 13dB
Ð
!
Mode No attenuator ÷10 attenuator ÷20 attenuator
Differential < ±400 mV < ±4 V < ± 8V
Common < ±16 V < ±42 V < ±42 V
Offset (Common) < ±1.6 V < ±16 V < ±32 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com