EasyManuals Logo

Agilent Technologies 1159A User Manual

Agilent Technologies 1159A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
36
1159A 1GHz Active Differential Probe
Calibration Test Record
Calibration Test Record
Agilent Technologies
1159A 1 GHz Differential Probe
Serial No.:_______________________
Certification Date:_________________
Tested By:_______________________
Recommended Test Interval: 1 Year
Recommended Date of Next Certification:_________
Certification Temperature:_____________________
_______________________________
_______________________________
Test Limit Minimum Results Limit Maximum
Accuracy at 1 kHz into 50
input of 54845 A
Min Gain 98% Max Gain 102%
Differential Mode Range
Att 1; Gain 1
Att 10; Gain 10
No Clipping
Pass
_____
_____
Fail
_____
_____
N/A
CMRR at 70 Hz 80 dB N/A
CMRR at 1 MHz 40 dB N/A
CMRR at 100 MHz 25 dB N/A
CMRR at 500 MHz 19 dB N/A
CMRR at 1 GHz 13 dB N/A