EasyManua.ls Logo

Agilent Technologies 1159A - 10:1 and 20:1 Attenuator Adjustment Procedure

Agilent Technologies 1159A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
The attenuator’s plastic covers may be removed. You must take care not to
damage the probe attenuator pins.
10:1 and 20:1 Attenuator Adjustment Procedure
Use the following steps to adjust the 10:1 or 20:1 attenuators.
1
Connect the function generator, probe and oscilloscope as shown.
Connect the + active probe input to the function generator output for these tests.
Connect the active input to ground with the probe ground socket.
2
Connect the 1159A Probe to CH1 of the oscilloscope.
3 Turn on the test equipment.
Allow 30 minutes for the probe to warm up.
Ð
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Related product manuals