5-13
Troubleshooting
To run the internal self-tests
To run the internal self-tests
Self Test performs a series of internal procedures to verify that the oscilloscope
and any attached module are operating properly.
It is recommended that you run the Self Test:
after experiencing abnormal operation
for additional information to better describe an oscilloscope failure
to verify proper operation after the oscilloscope has been repaired
Successfully passing Self Test does not guarantee 100% of the oscilloscope’s
functionality. Self Test is designed to provide an 80% confidence level that the
oscilloscope is operating properly.
1
Press the Utility key, then press the Service softkey.
2 Begin the internal self tests by pressing the Self Test softkey.
a The oscilloscope will display the message “Running self tests” on the screen.
b The oscilloscope will then display one of the following messages on the
screen indicating the status of the self-tests:
Self Tests Passed.
Self Tests Failed.
service.book Page 13 Wednesday, December 18, 2002 8:35 AM

Other manuals for Agilent Technologies 54622D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54622D and is the answer not in the manual?

Agilent Technologies 54622D Specifications

General IconGeneral
BrandAgilent Technologies
Model54622D
CategoryTest Equipment
LanguageEnglish

Related product manuals