EasyManua.ls Logo

Agilent Technologies 6060B - Page 13

Agilent Technologies 6060B
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
5. Measure the voltage across the monitor resistor and across the input terminals, then calculate the Electronic Load
resistance. The result should be between:
6060B: 0.0416 and 0.0584 ohms.
6063B: 0.792 and 1.208 ohms.
c. Check the middle ohms range as follows:
1. Press:
6060B:
Range 10 Enter, thenRES 30Enter
.
6063B:
Range 25 Enter, thenRES 500Enter
.
2. Set power supply for:
6060B: 10.9V and 15A.
6063B: 44V and 4A.
3. Measure the voltage across the monitor resistor and across the input terminals, then calculate the Electronic Load
resistance. The result should be between:
6060B: 24.1 and 39.6 ohms.
6063B: 433 and 590 ohms.
4. Then press:
6060B:
RES 1 Enter
.
6063B:
RES 24Enter
.
5. Measure the voltage across the monitor resistor and across the input terminals, then calculate the Electronic Load
resistance. The result should be between:
6060B: 0.989 and 1.011 ohms.
6063B: 23.75 and 24.25 ohms.
d. Check the high ohms range as follows:
1. Press:
6060B:
Range 1001 Enter, thenRES 120Enter
.
6063B:
RES 120 Enter, thenRES 2000Enter
.
2. Set power source for:
6060B: 60V and 6A.
6063B: 240V and 2A.
3. Measure the voltage across the monitor resistor and across the input terminals, then calculate the Electronic Load
resistance. Calculation should be between:
6060B: 61.1 and 3243 ohms.
6063B: 1247 and 5037 ohms.
4. Then press:
6060B:
RES 12Enter
.
6063B:
RES 240Enter
.
5. Measure the voltage across the monitor resistor and across the input terminals, then calculate the Electronic Load
resistance. The result should be between:
6060B: 10.9 and 13.3 ohms.
6063B: 223 and 259 ohms.