64 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
This page intentionally left blank.
Figure 31 Extract Chromatogram dialog box for Chloramphenicol (Negative mode)
Figure 32 Extract Chromatogram dialog box for Reserpine (Positive mode)
d For 6420/6460/6470, smooth the extracted chromatogram by setting
Smoothing Function to Quadratic/Cubic Savitzky-Golay, and Function
width to 9 points.
For 6410/6430/6490/6495, no smoothing is used to calculate S/N.