Other Setup Tasks 5
To confirm the collision cell gas flow rate (6470A and 6495A/B only)
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 95
7 For 6495A:
a In the Manual Tune tab, set Scan Type to MRM.
b In the MS1 tab, set SIM Mass to 322 amu.
Figure 43 Ion Source and MS1 settings
c In the MS2 tab, set SIM Mass to 290 amu.
Figure 44 MS2 settings
d In the Acquisition tab, set Dwell time to 200 ms.