EasyManuals Logo

Agilent Technologies 6813B User Manual

Agilent Technologies 6813B
91 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
B - Verification and Calibration
66
Table B-1. Equipment Required
Equipment Characteristics Recommended Model
Digital Voltmeter
Resolution: 10 nV @ 1 V
Readout: 8.5 digits
Accuracy: >20 ppm
Agilent 3458A
Current Monitor
1
0.01
,
±
200 ppm, 10 Watts
Guildline 7320/0.01
Ratio Transformer
2
30:1 ratio, 50 ppm, 45 Hz to 1 kHz
Load Resistor
20
, 10 A, 1800 Watts min.
Impedance Resistor
1
, 100 Watts min.
GPIB Controller
Full GPIB capabilities HP Series 200/300 or equivalent
1
The 4- terminal current shunt is used to eliminate output current measurement error caused by voltage drops in the
load leads and connections. It has special current-monitoring terminals inside the load connection terminals. Connect
the voltmeter directly to these current-monitoring terminals.
2
A ratio transformer is required only when verifying output voltage readback to MIL-STD-45662A 4:1 test
equipment ratio requirements.
Test Setup
Figure B-1 shows the setup for the tests. Be certain to use load leads of sufficient wire gauge to carry the
full output current (see Chapter 2).
Figure B-1. Verification & Calibration Test Setup
COM
R
Rs
L
Rs = Current Monitor resistor ( 0.01 ohms )
S1
S1 = Switch is for convenience (not required).
1
30
DMM
COM
01
SENSE
R = Load resistor for CC test (20 ohms)
01
L
Rz = Impedance resistor ( 1 ohm )
R
Z
DMM
Agilent 6811B/6812B/6813B
Agilent 3458A
Agilent 3458A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6813B and is the answer not in the manual?

Agilent Technologies 6813B Specifications

General IconGeneral
BrandAgilent Technologies
Model6813B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals