EasyManuals Logo

Agilent Technologies 8360 B Series Troubleshooting Guide

Agilent Technologies 8360 B Series
670 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
ILT.111
If a password is set, you must know it to do this step.
View calibration constant #463 (hardware conguration):
1. Press
4
PRESET
54
SERVICE
5
.
2.
Select
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adjust Menu
.
3. If
ENTER PASSWORD: 0
appears on the top line of the
display:
a. Enter the password.
b. Press
4
ENTER
5
.
c.
Select
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Adjust Menu
.
4.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Calib Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Select Cal
.
5. Press
4
463
54
ENTER
5
.
6. Note the value of calibration constants #463 and #464.
7. Note the options listed on the rear panel of the
instrument.
8. Note the model number of the instrument.
9. Using the options listed on the rear panel, the model
number, and the table below, calculate the hardware
conguration calibration constants (see the following
examples).
1a-8 Instrument Level Troubleshooting Agilent 8360

Other manuals for Agilent Technologies 8360 B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8360 B Series and is the answer not in the manual?

Agilent Technologies 8360 B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals