EasyManuals Logo

Agilent Technologies 8591E User Manual

Agilent Technologies 8591E
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
100 Chapter2
Making Adjustments
15. Frequency Response of the 8590L, 8591C, and 8591E
15.On the analyzer, press FREQUENCY and (step up), to step through
the remaining frequencies listed in Column 1 of Table 2-6. At each
new frequency repeat step 13 to step 15, entering the power sensor
Cal Factor into the measuring receiver as indicated in Column 3 of
Table 2-6.
16.On the synthesized sweeper, press CW and STEP UP.
Frequency Response Error At 4 MHz
17.Using a cable, connect the frequency synthesizer directly to the
INPUT 50. See Figure 2-29. For 75 inputs use a 75 cable to
connect the frequency synthesizer 75 OUTPUT to the INPUT 75
of the analyzer. Set the frequency synthesizer 50–75 switch to the
75 position. See Figure 2-29.
Figure 2-29 Frequency Response for 4 MHz Setup
CAUTION Use only 75 cables, connectors, or adapters on instruments equipped
with 75 inputs or damage to the input connectors will occur.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8591E and is the answer not in the manual?

Agilent Technologies 8591E Specifications

General IconGeneral
BrandAgilent Technologies
Model8591E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals