Chapter 2 2-135
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
33.Connect the equipment as shown in Figure 2-77.
Figure 2-77 Transmission Calibration Test Setup
34.Press
.
Load Match (Forward) Calibration Coefficient
35.Press .
36.When the test is done, press .
37.Using the front panel knob, locate the maximum value of the data trace for the 300 kHz
to 1.3 GHz frequency range.
38.Write the maximum value on the performance test record.
39.Repeat the previous two steps for the other frequency ranges listed on the performance
test record.
Transmission Tracking (Forward) Calibration Coefficient
40.Press .
41.When the analyzer finishes the test, repeat steps 37 through 39.
In Case of Difficulty
1. Check the condition of the connectors and cables.
TRANSMISSION
DO BOTH FWD THRUS
STANDARDS DONE
DONE FWD ENH RESP
System
SERVICE MENU
TESTS
36 x1
EXECUTE TEST
Marker
MARKER 1
System
SERVICE MENU
TESTS
37 x1
EXECUTE TEST

Table of Contents

Other manuals for Agilent Technologies 8753ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ES and is the answer not in the manual?

Agilent Technologies 8753ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals