2.2.1 Check condition of argon gas line
(1) Purge Ar line.
(2) Do leak check on Ar line.
(3) Check specification of Ar gas to have 99.99% purity (In case cylinder has been replaced).
(4) Try alternative source of Ar gas (In case cylinder has been replaced and any alternatives
are available)
(5) Does this fix the problem?
Yes No