Agilent ICP-MS
Interactive Troubleshooting tool for
Plasma ignition problem
Next >
Rev. B
2017, January, 20
Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ICP-MS Series and is the answer not in the manual?

Agilent Technologies ICP-MS Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelICP-MS Series
CategoryMeasuring Instruments
LanguageEnglish

Summary

Water and Gas Flow Troubleshooting

1.1 Check Water flow and Gas flow errors

Verify instrument status for specific error codes related to water and gas flow.

1.2 Check water and argon gas

Inspect cooling water and argon gas supply valves and levels for proper operation.

Argon Gas, Torch, and System Components

2.3 Torch Ignition Test

Temporarily block the torch end to test plasma ignition capability.

2.4 Torch Box Component Inspection and Issues

Inspect the torch box and its components for damage, incorrect assembly, and related issues.

Photo Sensor Cable Diagnostics

3.1 Photo Sensor Cable Check

Inspect the photo sensor cable for correct installation and signs of damage.

Vacuum System and Cone Checks

4.1 IF/BK Vacuum Pressure Errors

Check IF/BK vacuum pressure errors by reviewing the instrument status pane for specific error numbers.

Spray Chamber and Water Flow Issues