EasyManuals Logo

Agilent Technologies DSO3062A User Manual

Agilent Technologies DSO3062A
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
Service
DC Gain Accuracy Test
4-8
Figure 4-1
4 Select the Default Setup item in the Save/Recall menu.
5 Press the
Acquire front panel button.
6 Select the mode item in the Acquire menu until Average appears.
7 Select the
Averages item in the Acquire menu until 256 appears.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies DSO3062A and is the answer not in the manual?

Agilent Technologies DSO3062A Specifications

General IconGeneral
BrandAgilent Technologies
ModelDSO3062A
CategoryTest Equipment
LanguageEnglish

Related product manuals