Service
DC Gain Accuracy Test
4-12
27 Calculate the DC Gain using the following expression and record this
value in the DC Gain Test section of the Performance Test Record:
DCGain =
V
out
V
in
--------------
=
V
scope+
V
scope-
V
DMM+
V
DMM-
-----------------------------------------