ESG Family Signal Generators Using Functions
Creating a Step Sweep and a List Sweep
User’s Guide 2-9
Creating a Step Sweep and a List Sweep
Using this procedure, you will learn two ways to set up the signal generator to sweep a
defined set of points. You will create a step sweep and then you will use these points as the
basis for a new list sweep.
In the first procedure, you will create a step sweep with the following ten, equally spaced
points:
frequency range from 525 MHz to 600 MHz
power level from
20 dBm to 0 dBm
dwell time 500 ms at each point
In the second procedure, you will take the step sweep points and edit several points to
change the sweep information.
Configuring a Step Sweep
1. Preset the signal generator to the factory-defined instrument state.
2. Press the front panel
Sweep/List key. The first level of sweep softkeys is displayed. Toggle
the
Sweep Type List Step softkey to Step.
3. Press the
Configure Step Sweep softkey. Another menu is displayed with softkeys that you
will use to create the sweep points.
4. Change the start frequency of the step sweep. Press the
Freq Start softkey. Enter
525 MHz using the numeric keypad and pressing the
MHz terminator softkey.
5. Change the stop frequency of the step sweep. Press the
Freq Stop softkey. Enter 600 MHz
using the numeric keypad and pressing the
MHz terminator softkey.
6. Set the power level for the start of the step sweep. Press the
Ampl Start softkey. Enter
20 dBm using the numeric keypad and pressing the
dBm terminator softkey.
7. Set the power level for the end of the step sweep. Press the
Ampl Stop softkey. Enter
0 dBm using the numeric keypad and pressing the
dBm terminator softkey.
8. Set the number of sweep points. Press the
# Points softkey. Enter 8 by rotating the front
panel knob until the number 8 is displayed.
9. Set the dwell time at each point. Press the
Step Dwell softkey, then enter 500 ms using
the numeric keypad and pressing the
msec terminator softkey.

Other manuals for Agilent Technologies E4421B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4421B and is the answer not in the manual?

Agilent Technologies E4421B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4421B
CategoryPortable Generator
LanguageEnglish

Related product manuals