Battery Management Functionality
E7495A/B Manual Verification
Running the Battery Management Functional Test
NOTE When connecting or removing the Ext DC Power, the E7495A/B Test Set
can take up to 20 seconds for the Battery Status Indicators to update.
1. External DC Power Test
Setup
Install battery into Battery Position #2 (Closest to the back of
the E7495A/B case)
Plug the Ext DC power into the E7495A/B
Turn the Ext DC power supply’s switch to on
Turn the E7495A/B on and allow it to fully power-up
Test Results
On the Front Panel Display, verify that the External DC icon
(Plug icon) is present on the lower right hand portion of the
display.
Verify that the Battery Icons on the Front Panel Display
accurately represent the condition of the installed battery.