17 Utility
17.4 ForcePG12: PG Output1/Output2 (A.01.20)
[Supported Analyzer]
B1500A, B1505A, 4155B, 4155C, 4156B, 4156C
[Description]
Sets Output1 and Output2 of Agilent 81110A Pulse Generator, and triggers it. Pulse leading/trailing edge
transition time can be set.
Execute ResetPG to stop pulse output before the specified pulses are applied.
[Required Modules and Accessories]
Agilent 81110A Pulse Generator (2-output, PGU1 and PGU2)
GPIB cable
[Test Parameters]
Address: GPIB address of the Agilent 81110A Pulse Generator
Period1: Output1 pulse period [s]
Delay1: Output1 delay time [s]
Dcyc1: Output1 duty cycle [%]
LeadTime1: Output1 pulse leading edge transition time [s]
TrailTime1: Output1 pulse trailing edge transition time [s]
Level1: Output1 pulse high level [V]
Base1: Output1 pulse low level [V]
ExtImp1: Output1 load impedance [ohm]
NoOfPulse12: Number of output pulses (1 to 65536)
Period2: Output2 pulse period [s]
Delay2: Output2 delay time [s]
Dcyc2: Output2 duty cycle [%]
LeadTime2: Output2 pulse leading edge transition time [s]
TrailTime2: Output2 pulse trailing edge transition time [s]
Level2: Output2 pulse high level [V]
Base2: Output2 pulse low level [V]
ExtImp2: Output2 load impedance [ohm]
Agilent EasyEXPERT Application Library Reference, Edition 8
17-6