Maintaining the Instrument
Changing the Sample Loop
93
Figure 20. Removing the manifold cover and insulation
3. Use a 7-mm wrench to loosen the two nuts connecting the sample loop to
the six–port valve. See Figure 21.
Loosen
Manifold cover
and insulation
(Front cover not shown
for clarity.)